OM0303
Rev 5 July/2021
Page 41 of 108
Perform decontamination procedure per NSF/ANSI 49, Annex I-2 (formerly Annex G).
If the cabinet has been used to prepare antineoplastic drugs, (chemotherapy), or other toxic chemicals,
decontamination of the cabinet
cannot
be accomplished by the above procedure. It is recommended that the
following protective measures be taken:
1.
Gloves
Gloves must be worn. Care must be taken not to cut, puncture, or tear the gloves. No one glove material
is impervious to all CYTAs; disposable surgical or polyvinyl chloride (PVC) gloves provide substantial but
not complete protection. PVC gloves probably are more protective than surgical gloves, but they are
stiffer and less tactile. Gloves should be discarded after each use. Gloves should be tucked into the cuffs
of the gown. Double gloving should be considered.
2.
Face & Eye Protection
A disposable dust and mist respirator and either a plastic face shield (preferred) or chemical splash
goggles must be worn. The face shield or goggles should be wiped clean with a suitable tissue and water
after each use.
3.
Gowns
A protective garment must be worn. The garment should be made of lint-free, low-permeability fabric
and must have a closed front, long sleeves, and elastic or knit closed cuffs. Tyvek¹ isolation gowns are one
example of an acceptable garment. The garment must be worn outside the work area. Disposable gowns
are preferred over reusable. Front-buttoned coats are not recommended.
4.
Hair & Shoe Covers
Disposable hair and shoe covers should be worn.
5.
Motion
Slow and deliberate motions are necessary when working in the interior of the cabinet, in order to
minimize the generation of particulates.
Please consult with Nuaire, Inc. about any unique contamination problems.
Normally, no preventive maintenance is required on the interior of the cabinet (i.e., the area behind the access
panel containing the HEPA filters and motor (blower assembly). All required adjustments in order to maintain
proper cabinet airflows are external to the cabinet interior. The motor is lubricated for life and is thermally
protected with automatic reset.
¹ Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply distributor
Содержание LabGard NU-581-400E
Страница 7: ...OM0303 Rev 5 July 2021 Page 7 of 108 ...
Страница 9: ...OM0303 Rev 5 July 2021 Page 9 of 108 ...
Страница 10: ...OM0303 Rev 5 July 2021 Page 10 of 108 ...
Страница 11: ...OM0303 Rev 5 July 2021 Page 11 of 108 ...
Страница 15: ...OM0303 Rev 5 July 2021 Page 15 of 108 ...
Страница 30: ...OM0303 Rev 5 July 2021 Page 30 of 108 Set New Password Retype new password ...
Страница 43: ...OM0303 Rev 5 July 2021 Page 43 of 108 ...
Страница 101: ...OM0303 Rev 5 July 2021 Page 101 of 108 ...
Страница 102: ...OM0303 Rev 5 July 2021 Page 102 of 108 ...
Страница 103: ...OM0303 Rev 5 July 2021 Page 103 of 108 ...
Страница 104: ...OM0303 Rev 5 July 2021 Page 104 of 108 ...
Страница 105: ...OM0303 Rev 5 July 2021 Page 105 of 108 ...
Страница 106: ...OM0303 Rev 5 July 2021 Page 106 of 108 ...
Страница 107: ...OM0303 Rev 5 July 2021 Page 107 of 108 ...
Страница 108: ...OM0303 Rev 5 July 2021 Page 108 of 108 ...