OM0303
Rev 5 July/2021
Page 57 of 108
7.5.2.2 Setpoints/Limits
The airflow setpoints and alarm limits may also be verified or altered. Typically, these default values are
factory set based on the cabinet type, model and size as previously discussed. However, they may be altered
in special cases for modified cabinets. The setpoint establishes the airflow values that are to be maintained.
The high low limits establish the alarm boundaries from the nominal setpoint. The default values have been
established based upon the performance specifications and cabinet component tolerances.
Entry into the setpoints/limits
functions requires a service password for entry. The default password is “9876”
Default values for NU-581E
•
Downflow setpoint - .30
•
Downflow high limit - .33
•
Downflow low limit - .27
•
Inflow setpoint - .53
•
Inflow high limit - .61
•
Inflow low limit
–
.46
Содержание LabGard NU-581-400E
Страница 7: ...OM0303 Rev 5 July 2021 Page 7 of 108 ...
Страница 9: ...OM0303 Rev 5 July 2021 Page 9 of 108 ...
Страница 10: ...OM0303 Rev 5 July 2021 Page 10 of 108 ...
Страница 11: ...OM0303 Rev 5 July 2021 Page 11 of 108 ...
Страница 15: ...OM0303 Rev 5 July 2021 Page 15 of 108 ...
Страница 30: ...OM0303 Rev 5 July 2021 Page 30 of 108 Set New Password Retype new password ...
Страница 43: ...OM0303 Rev 5 July 2021 Page 43 of 108 ...
Страница 101: ...OM0303 Rev 5 July 2021 Page 101 of 108 ...
Страница 102: ...OM0303 Rev 5 July 2021 Page 102 of 108 ...
Страница 103: ...OM0303 Rev 5 July 2021 Page 103 of 108 ...
Страница 104: ...OM0303 Rev 5 July 2021 Page 104 of 108 ...
Страница 105: ...OM0303 Rev 5 July 2021 Page 105 of 108 ...
Страница 106: ...OM0303 Rev 5 July 2021 Page 106 of 108 ...
Страница 107: ...OM0303 Rev 5 July 2021 Page 107 of 108 ...
Страница 108: ...OM0303 Rev 5 July 2021 Page 108 of 108 ...