OM0303
Rev 5 July/2021
Page 59 of 108
7.5.3.1 Downflow Calibration
Step 1:
•
Access Calibration/Service menu.
•
Enter Password
•
Press SENSOR SETUP menu item to access sensor calibration menu.
Step 2:
•
Press calibrate downflow sensor. If blower was not running, blower will automatically
turn on. Run for one minute or until downflow readings are steady.
Step 3:
•
Place a velometer in the cabinet workzone on the horizontal plane 4 inches (102mm)
above the bottom edge of the window. Spot check several points on the recommended
downflow velocity test grid found in table 7.0.
Step 4:
•
Press +/- to adjust supply duty cycle.
The objective of this spot check is to obtain the desired downflow average velocity as
close as possible to the stated goal of 60 LFPM (.30 m/s)..
DON'T SPEND MORE THAN 5 MINUTES SPOT CHECKING.
Содержание LabGard NU-581-400E
Страница 7: ...OM0303 Rev 5 July 2021 Page 7 of 108 ...
Страница 9: ...OM0303 Rev 5 July 2021 Page 9 of 108 ...
Страница 10: ...OM0303 Rev 5 July 2021 Page 10 of 108 ...
Страница 11: ...OM0303 Rev 5 July 2021 Page 11 of 108 ...
Страница 15: ...OM0303 Rev 5 July 2021 Page 15 of 108 ...
Страница 30: ...OM0303 Rev 5 July 2021 Page 30 of 108 Set New Password Retype new password ...
Страница 43: ...OM0303 Rev 5 July 2021 Page 43 of 108 ...
Страница 101: ...OM0303 Rev 5 July 2021 Page 101 of 108 ...
Страница 102: ...OM0303 Rev 5 July 2021 Page 102 of 108 ...
Страница 103: ...OM0303 Rev 5 July 2021 Page 103 of 108 ...
Страница 104: ...OM0303 Rev 5 July 2021 Page 104 of 108 ...
Страница 105: ...OM0303 Rev 5 July 2021 Page 105 of 108 ...
Страница 106: ...OM0303 Rev 5 July 2021 Page 106 of 108 ...
Страница 107: ...OM0303 Rev 5 July 2021 Page 107 of 108 ...
Страница 108: ...OM0303 Rev 5 July 2021 Page 108 of 108 ...