All interlocks are enable/ disabled by a user through the interlock Features screen
o
UV window interlock:
This parameter allows for the selection of the window closed switch to be interlocked with the UV light
option. When the window interlock is on, the window must be closed for the UV light to operate.
When the window interlock is off the UV light can be turned on regardless of the window position.
NOTE
: In addition to the
TOUCHLINK
system UV window interlock there is a double redundant UV
window
interlock relay. To override the UV window interlock for service purposes only, both interlocks must be
changed through the
TOUCHLINK
system and shorting the relay connection. (See electrical schematic
for reference)
Содержание LabGard NU-581-400E
Страница 7: ...OM0303 Rev 5 July 2021 Page 7 of 108 ...
Страница 9: ...OM0303 Rev 5 July 2021 Page 9 of 108 ...
Страница 10: ...OM0303 Rev 5 July 2021 Page 10 of 108 ...
Страница 11: ...OM0303 Rev 5 July 2021 Page 11 of 108 ...
Страница 15: ...OM0303 Rev 5 July 2021 Page 15 of 108 ...
Страница 30: ...OM0303 Rev 5 July 2021 Page 30 of 108 Set New Password Retype new password ...
Страница 43: ...OM0303 Rev 5 July 2021 Page 43 of 108 ...
Страница 101: ...OM0303 Rev 5 July 2021 Page 101 of 108 ...
Страница 102: ...OM0303 Rev 5 July 2021 Page 102 of 108 ...
Страница 103: ...OM0303 Rev 5 July 2021 Page 103 of 108 ...
Страница 104: ...OM0303 Rev 5 July 2021 Page 104 of 108 ...
Страница 105: ...OM0303 Rev 5 July 2021 Page 105 of 108 ...
Страница 106: ...OM0303 Rev 5 July 2021 Page 106 of 108 ...
Страница 107: ...OM0303 Rev 5 July 2021 Page 107 of 108 ...
Страница 108: ...OM0303 Rev 5 July 2021 Page 108 of 108 ...