PEEP
Positive end-expiratory
pressure.
Setting
parameter
Description
Plow
Plow is the low pressure level at which the patient can
breathe
spontaneously.
Phigh
Phigh is the high pressure level at which the patient can
spontaneously
breathe and is an absolute
value.
Pinsp
It is a relative value of the pressure, relative to
PEEP.
Psupp
Pressure
support
level
in
pressure
control
mode.
It
is
a
relative
value
relative
to
PEEP or
Plow.
Tslope
Controls pressure rise slope in pressure
mode.
Tpause(%)
Percent of gas delivery pause time in inspiratory time within
the
inspiratory phase.
Rate
The number of mechanically controlled breaths delivered to the
patient in
one minute.
Thigh
Thigh is the time that the ventilator will hold the high pressure
level.
Tlow
Tlow is the time that the ventilator will hold the low pressure
level.
Tinsp
Inspiration Time in one breathing
cycle.
Trigger
Pressure trigger and flow trigger included. When the trigger level is
detected,
the ventilator starts to enter the inspiratory phase.
Exp%
Inspiratory termination level. The ventilator is switched to the
expiratory phase when the inspiratory flow drops to
peak flow*Exp
%
.
Apnea ventilation
Turn on or turn off apnea ventilation
function.
Apnea Rate
Breathing frequency set in apnea ventilation
mode.
Apnea Vt
It is delivered tidal volume in apnea ventilation when volume
mode is
selected for apnea
ventilation.
Apnea
Tinsp
Inspiration time set in apnea ventilation
mode.
Sigh
Turn on or turn off sigh
function.
Interval
It is the setting value of number of cycles of every group of
sigh
ventilation.
Delta Vt
It is intermittent vt augmentation, added during the sigh
cycle.
Delta Pinsp
It is intermittent pinsp augmentation, added during the sigh
cycle.
Disable ATRC
Turn on or turn off ATRC
function.
ET
Tube
Initiate ATRC function for ET
tube.
Trach
Tube
Initiate ATRC function for Trach
tube.
Tube
I.D.
It refers to the diameter of tracheal or ET
tube.
Compliance
compensate Com
Turn on or turn off compliance compensate
function.
Содержание Crius V6
Страница 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Страница 15: ...15 182 FOR YOUR NOTES...
Страница 26: ...26 182 FOR YOUR NOTES...
Страница 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Страница 32: ...32 182 FOR YOUR NOTES...
Страница 44: ...44 182 FOR YOUR NOTES...
Страница 56: ...56 182 FOR YOUR NOTES...
Страница 88: ...88 182 FOR YOUR NOTES...
Страница 100: ...100 182 FOR YOUR NOTES...
Страница 125: ...125 182 FOR YOUR NOTES...
Страница 149: ...149 182 FOR YOUR NOTES...
Страница 156: ...156 182 FOR YOUR NOTES...