Check the patient
type.
2.
Check the alarm
limits.
Apnea
CO
2
M
The time of failure to detect respiration by
the
CO
2
module exceeds Apnea
Tinsp.
1.
Check the
patient.
2.
Check apnea time
setup.
3.
Check the connections of CO
2
module sampling
device.
SPO2 Too High
M
SpO2 value is higher than the upper alarm limit or lower
than the lower alarm limit. Check the patient's physiological
condition, and check if the patient category and alarm limit
settings are appropriate for the patient.
SPO2 Too Low
M
PR Too High
M
SpO2 value is higher than the upper alarm limit or lower
than the lower alarm limit. Check the patient's physiological
condition, and check if the patient category and alarm limit
settings are appropriate for the patient.
PR Too Low
M
10.8.2
Technical alarm
Technical Alarm Table
Alarm message
Level
Cause and
action
Battery Low
M
The remaining battery power is lower than a
threshold.
Connect to the external power
supply.
The system will shut down
H
Battery power is depleted. The system will shut down in
a few minutes.
Connect to the external power supply immediately.
Battery Disconnect
L
Equipment is not installed the battery
Please install the battery, so as not to power off the
external power supply, battery-powered.
AC Disconnect
L
The device is not connected to an external power supply
Please access the external power supply.
MERCURY Comm.
Malfunction
H
Breathing communication failure.
Please check if the ventilator communication USB
interface cable is loose.
Apnea Ventilation
M
Current mode ventilation patients without inspiratory
triggering ability.
Please check the patient.
Insp. Vt Limit Reached
M
Exhaled air volume exceeds the set tidal volume.
1. Check the patient.
2. Check the ventilation parameter settings.
3. Check the pressure alarm high limit.
Insp. Flow Trigger
L
Inspiratory flow rate trigger.
1. Check the patient.
2. Check ventilation settings.
Содержание Crius V6
Страница 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Страница 15: ...15 182 FOR YOUR NOTES...
Страница 26: ...26 182 FOR YOUR NOTES...
Страница 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Страница 32: ...32 182 FOR YOUR NOTES...
Страница 44: ...44 182 FOR YOUR NOTES...
Страница 56: ...56 182 FOR YOUR NOTES...
Страница 88: ...88 182 FOR YOUR NOTES...
Страница 100: ...100 182 FOR YOUR NOTES...
Страница 125: ...125 182 FOR YOUR NOTES...
Страница 149: ...149 182 FOR YOUR NOTES...
Страница 156: ...156 182 FOR YOUR NOTES...