182
Fig. 8-1 SpO2 Parameter Area
8.5
Setting SpO2
Select SpO2 parameter area or Pleth wave area → [SpO2 Setup] menu, which is shown
below. You can set SpO2 through [SpO2 Setup] menu.
Fig. 8-2 [SpO2 Setup] Menu
8.5.1
Setting Wave Speed
Select [Wave Speed] and set wave speed to [12.5mm/s] or [25mm/s]; the faster speed,
the smoother wave.
10.5.1
Setting Wave Mode
Select [Wave Mode], and set the wave drawing mode
[Scan]: Scan mode.
[Fill]: Fill mode.
10.5.2
Setting Average Time
SpO2 values displayed on the monitor are the results averaged from the data collected over
time. The shorter the average time, the faster the monitor responds when the patient's SpO2
value changes, but the measurement accuracy is lower. Conversely, the longer the average
Содержание Crius V6