73
/
182
1.
[
Disable ATRC
]: Switch off ATRC function
2.
[
ET Tube
]: Endotracheal tube
3.
[
Trach Tube
]:
Tracheal tube
4.
[
Tube I.D.
]: Diameter of tube
5.
[
Compensate
]:
Compensate portion
6.
[
Expiration
]:
Compensate of expiration
In VC-ACV mode,you can set the other function parameters
as required in Invasive ventilation
(this function is applied to all Invasive modes and the description is not repeated in the other
ventilation modes section in this manual.)
:
1.
[Compliance Compensate]
2.
[Exp.Speed]
6.5.4 PC-ACV
PC-ACV (Pressure Controlled– Assist Control Ventilation) is a ventilation mode in which the
ventilator controls the ventilation, but it can be influenced by the patient. The inspiration and
expiration are pressure controlled. The patient can breathe spontaneously during inspiration and
expiration.
The strokes are time cycled with a fixed inspiratory time and a fixed expiratory time, when the
patient does not breathe spontaneously. However, during expiration an inspiration can be
triggered by the patient. This can lead to an increase of the breathing frequency or respiratory
rate(RR).
Содержание Crius V6
Страница 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Страница 15: ...15 182 FOR YOUR NOTES...
Страница 26: ...26 182 FOR YOUR NOTES...
Страница 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Страница 32: ...32 182 FOR YOUR NOTES...
Страница 44: ...44 182 FOR YOUR NOTES...
Страница 56: ...56 182 FOR YOUR NOTES...
Страница 88: ...88 182 FOR YOUR NOTES...
Страница 100: ...100 182 FOR YOUR NOTES...
Страница 125: ...125 182 FOR YOUR NOTES...
Страница 149: ...149 182 FOR YOUR NOTES...
Страница 156: ...156 182 FOR YOUR NOTES...