Exp. Flow Trigger
L
Exhalation flow rate trigger.
1. Check the patient.
2. Check ventilation settings.
Pressure Limited
L
In volume mode or pressure mode when
ATRC
function is enabled, the pressure reaches Paw
high alarm
limit-5.
1.
Check the
patient.
2.
Check the ventilation parameter
setup.
3.
Check pressure high alarm
limit.
O
2
% Too
High
H
During O
2
therapy, the O
2
concentration is greater
than
the O
2
% high alarm limit for at least
30s.
1.
Check the ventilation parameter
setup.
2.
Check the alarm
limits.
3.
Check the O
2
supply.
4.
Calibrate the O
2
sensor.
O
2
% Too
Low
H
During O
2
therapy, the O
2
concentration is less
than
the O
2
% low alarm limit for at least 30s or is
less than 18
%.
1.
Check the ventilation parameter
setup.
2.
Check the O
2
supply.
3.
Calibrate the O
2
sensor.
PEEP Too
High
H
Monitored PEEP exceeds PEEP+5 cmH
2
O
within
any fully mechanical ventilation
cycle.
1.
Check the ventilation parameter
setup.
2.
Check the patient tubing for
occlusion.
PEEP Too
Low
M
Patient’s PEEP is less than the setting value to
a certain extent.
1.
Check the patient tubing for
leakage.
2.
Perform System Check to test the
leakage.
Tube Leak
L
Tube is
leaky.
1.
Check the patient tubing for
leakage.
2.
Perform System Check to test the
leakage
Hose Disconnected
H
Tube is
disconnected.
Re-connect the patient tubing.
Paw Senseor Malfunction
H
Pressure sensor failure.
1. Calibrate the pressure sensor.
2. Please contact your designated service representative.
No Available Pressure
Sensor
H
There is no reliable pressure sensor.
1. Check that the pressure sensor is in place.
2. Please calibrate the pressure sensor.
Please contact your designated service representative.
CO
2
Sampleline
Occlud
ed
L
Sampling line is faulty or occluded.
1. Check the sampling line for occlusion.
2. Replace the sampling line.
3. Replace the water trap.
O2 pressure is low or high-pressure O2 is not connected.
Содержание Crius V6
Страница 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Страница 15: ...15 182 FOR YOUR NOTES...
Страница 26: ...26 182 FOR YOUR NOTES...
Страница 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Страница 32: ...32 182 FOR YOUR NOTES...
Страница 44: ...44 182 FOR YOUR NOTES...
Страница 56: ...56 182 FOR YOUR NOTES...
Страница 88: ...88 182 FOR YOUR NOTES...
Страница 100: ...100 182 FOR YOUR NOTES...
Страница 125: ...125 182 FOR YOUR NOTES...
Страница 149: ...149 182 FOR YOUR NOTES...
Страница 156: ...156 182 FOR YOUR NOTES...