SPECIFICATION - ANCILLARIES
The Ancillaries connector contains inputs and outputs for connection to external equipment.
Pin Connections
Pin No
Function
1
D5 V (maximum current 5 mA)
2
External PRINT / RESET Input (Active High)
3
Low Limit Torque Output
4
Pass Limit Torque Output
5
High Limit Torque Output
6
Low Limit Angle Output
7
Pass Limit Angle Output
8
High Limit Angle Output
9
Digital 0 V
10
Tool Stop Output
11
Analogue Output
12
Analogue Output 2.5 V
13
Analogue Output 0 V reference (Do not connect to a noisy electrical ground)
14
External Snug Trigger Input
15
Not Used
External Print / Reset Input
Pins 1 & 2 are intended for use as an EXTERNAL PRINT / RESET:-
The switch must remain active for at least 200 mS. Screened cable is recommended.
Limit Outputs
Pins 3, 4, 5, 6, 7 & 8 are buffered logic outputs intended for Go/No Go control of external equipment.
The limit outputs are referenced to Pin 9 (Digital 0 V).
All limit outputs are active HIGH and are updated every 10 mS
The limit outputs change exactly with increasing torque, and at 0.5% of transducer capacity below the limit
with decreasing torque. This eliminates the logic lines oscillating.
The Limit output current is, High = -0.8 mA, Low = 16 mA (not for direct control of relays).
Tool Stop Output
Tool Stop is used to provide a stop signal for an external pneumatic, hydraulic or electric tool. When the
measured torque goes above the Tool Stop value (set in the Target Setup) this pin goes HIGH (5 V) with
reference to pin 9.