12
Noise
Suppression
Allows the user to disable noise suppression in order to capture and analyse extremely fast
signals.
The data supplied in the T-Box™ 2 calibration certificate is obtained with noise
suppression enabled. Disabling it may result in wider uncertainties due to noise.
The password to disable Noise Suppression is: transients
Filter
Frequency
Selects an AnB module’s ADC filter frequency. Signals with frequencies above the current
setting are reduced. Increase the filter frequency for signals with rapid transients.
Serial
Output
Enable or disable serial port output of readings or live data streams for this specific AnB
module (requires serial port output enabled in settings > serial ports)
Set
This will set the current measurement mode based on the settings above
Cancel
Cancel mode setting; any changes will not be saved
NOTE:
Not all settings are available in all modes; some are mode-specific.
TIP: If you want to capture graphs, bear in mind the AnB mode settings will affect the capture, so it
is important to consider the mode before capture begins. If you need to check or change the
mode, access Mode Selection from the Measurement Screen.
The mode effects on graphing are summarised below:
Track:
Best for slow-moving signals and when maximum noise rejection is required.
Click:
Best for graphing the behaviour of click-type setting tools such as
ISO 6789:2017 Type II Class A tools.
Dial:
Best for any other scenario than those outlined above. Probably, the most
common all-purpose mode to use for graphing would be Dial mode.
Stall/Screwdriver/Hydraulic: There is no difference in graphing performance between these modes and
Dial mode.