Sensor type
Setting and choosing the several sensor types is the same procedure as if in
case of using only one flow velocity sensor.
Mounting place
When using multiple sensors enter the mounting height for each sensor here.
Please observe that sensor 1 operates as a guide, i.e. all level readings refer to
this sensor. This is why this sensor has to be installed at the lowest point of the
application (see Fig. 9-61)
Note
If several flow velocity sensors are used and one of them operates with ultra-
sonic level measurement from bottom up or pressure measurement cell, this
sensor is always sensor 1 within the sequence and hence has to be installed
on the lowest point of the application. One surface radar Type OFR can be
installed per measurement place only.
Fig. 9-62
Sensor Assignment if using multiple v-sensors
Note
If sensors 2 and 3 are located higher than sensor 1, this level difference shall
be entered as height "h" under the menu point mounting place. In order to
obtain a total result the velocity will not be referred to below this level.