Fig. 9-63
Individual value assignment of flow velocity sensors
Interval d
(Distance)
Distance "d" is the distance to the profile centreline. This parameter currently
has no function and hence is not used for calculations. Any entries here will not
influence the measurement result.
Percent
This parameter defines the share of the sensor regarding the total result. The
following relation applies for the percentage of the individual flow velocities:
x% + y% + z%
=
x%
or
y%
or
z%
100%
share x
share y
share z
x%, (y%), (z%) = percentage set sensor 1, (2), (3)
share x, (y), (z) = share of sensor in total velocity
9.5.4 Parameter Menu "Analog Inputs “
Fig. 9-64
Analog inputs – submenu
Various numbers of analog inputs are available depending on the type of trans-
mitter, such as in case of transmitter type >S4<:
-
1 analog input (galvanically isolated) for 2-wire sensors
-
1 additional analog input for external level sensor connection
If using transmitter type >M4<:
-
1 analog input (galvanically isolated) for 2-wire sensors
-
4 additional analog inputs for external level sensor connection as well as for
external setpoints or analog data storage.