Fig. 9-78
Setting the set point type
Note
Analog input 4 is the dedicated hardware input for analog setpoint signals.
This assignment cannot be modified.
Parameter:
Internal set point:
-
define the internal setpoint in indicated unit
External set point:
-
name (it is not necessary to enter a name, since it currently is used for de-
vice-internal purposes only)
-
measurement range of external setpoint, select from 0/4-20 mA.
0-5/10 V can be set by NIVUS service personnel only.
-
set point input linearisation (normally >2< is going to be entered as number
of breakpoints. Set the setpoint start (=0) to 0/4-20 mA and the setpoint stop
to 20 mA subsequently. The input area can be linearised as well.)
Offset:
-
this values is going to be added to the external setpoint. Negative values are
allowed to be set as well.
Cable break monitoring is being executed if the external setpoint has been set to
a range of 4-20 mA. It is going to be switched over to the internal setpoint as
soon as a cable break has been detected (default setting = 100 l/s).
Relays
The logic functions of both output relays can be modified using this menu.
Fig. 9-79
Assignment of relay functions
Control CLOSED:
here you can enter the internal name (not required) as well
as the logic function of relay 4 (normally closed or normally open). Select by
pressing >ALT<.
Содержание OCM PRO CF
Страница 179: ...20 Certificates and Approvals...
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 183: ......
Страница 184: ......
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 188: ......
Страница 189: ......
Страница 190: ......
Страница 191: ......
Страница 192: ......