Chapter 10 User interfaces
Note:
Example to explain the filter time in conjunction with the ext. warmup time
Basic setting
Record interval
5min.
Measurement cycle
1min.
Output channels
Ext. warmup time
1sec.
Measurement channels -
>Config.
Filter time
500ms
Output on the device
Sensor supply
Explanation:
The sensor supply is always activated 1sec before expiry of the measurement cycle. The filter
time starts 500ms before expiry of the measurement cycle, which ensures that a valid value is available at the
time of a measurement. However, this also means that the filter time must be taken into consideration when
selecting the ext. warmup time. In the current example, the sensor connected to the input must at the latest
supply stable values 500ms following activation of the sensor supply. Otherwise, invalid values would be taken
into consideration while averaging during the filter time resulting in the measurement value being distorted.
NivuLog Easy V3 - Rev. 03 as of 24.10.2018
91 of 141
Содержание NivuLog Easy V3
Страница 2: ...2 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 8: ...8 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 16: ...16 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 20: ...20 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 42: ...42 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 68: ...68 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 110: ...110 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 112: ...112 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 118: ...118 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 120: ...120 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 130: ...130 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 138: ...138 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...
Страница 140: ...140 of 141 NivuLog Easy V3 Rev 03 as of 24 10 2018...