Parameter Setting
NFP - rev. 04 / 23.05.2018
page 67
>up< and >down< after the PIN code “2718” (safety function) has
been entered. Furthermore it is possible to directly enter the desired
simulation value by pressing ENTER. A maximum output current of
21 mA can be simulated (see Fig. 31-4/2).
Fig. 31-4
Simulation of analog outputs
31.3 CAL Menu “Relays”
>Relays<
Selecting the >Relay Outputs< option requires to enter the system
PIN once again to make sure that simulations can be carried out by
authorised personnel only.
Select the desired relays to simulate by using the >up< or >down<
arrow keys. Use ENTER to directly energise or de-energise the cho-
sen relay. Energised relays will de-energise as soon as you exit the
menu.
Fig. 31-5
Relay simulation