Instruction Manual
NIVUS Full Pipe
page 24
NFP - rev. 04 / 23.05.2018
20.3 Hints on how to avoid electrostatic discharge (ESD)
When connecting the NFP the following warnings and hints shall be observed right as the
warnings and hints found in the according chapters on installation.
WARNING
Disconnect the unit from mains power
Disconnect the instrument from mains power before you begin to carry out maintenance,
cleaning and/or repair works (expert personnel only).
Disregarding may lead to electric shocks.
The sensitive electronic components inside of the instrument may be damaged by static elec-
tricity which may impair the instrument’s performance or even lead to instrument failure. The
manufacturer recommends the following steps to avoid equipment damage due to electrostat-
ic discharge:
•
Discharge static electricity from your body before touching the instrument’s electronic
components such as circuit boards and the components installed on the boards. To do
so touch a grounded metal surface such as the unit’s enclosure frame or a metal pipe.
•
Avoid unnecessary movements to reduce the risk of building up static electricity.
•
Transport statically sensitive components in antistatic containers or packing materials.
•
To discharge your body and to stay free of static electricity wear an antistatic wristband
grounded through a cable.
•
Only touch components that are sensitive to electric charges in an antistatic working ar-
ea. If possible, use antistatic mats and work pads.
20.4 Transmitter Installation
Front panel
It is not allowed to remove the front panel.
Tightness of the terminal housing
Water or dirt must not leak into the terminal housing. Seal the housing with the supplied lid
and both screws respectively.
Mounting Wall Mount Enclosure
Ensure proper installation.
The most simple way to install a wall mount enclosure is to fasten a DIN rail with a length of
210 mm (8.3 in.) and then to snap-on the enclosure.
It is possible to install the enclosure by using three screws as well. Use a pan head screw
with a head diameter of 5.5...8.0 mm (0.22...0.32 in.) for this. This screw must be screwed in-
to the mounting plate protruding 4 mm (0.16 in.). Then hang the enclosure on the screw and
additionally fix it with two more screws from the terminal clamp housing. Make sure to screw
Содержание NFP
Страница 82: ...Instruction Manual NIVUS Full Pipe page 82 NFP rev 04 23 05 2018 Approvals and certificates...
Страница 83: ...Approvals and certificates NFP rev 04 23 05 2018 page 83...
Страница 84: ...Instruction Manual NIVUS Full Pipe page 84 NFP rev 04 23 05 2018...
Страница 85: ...Approvals and certificates NFP rev 04 23 05 2018 page 85...
Страница 86: ...Instruction Manual NIVUS Full Pipe page 86 NFP rev 04 23 05 2018...
Страница 87: ...Approvals and certificates NFP rev 04 23 05 2018 page 87...