Instruction Manual
NIVUS Full Pipe
page 48
NFP - rev. 04 / 23.05.2018
The >right< or >left< arrow keys can be used to move the cursor
even after the cursor reaches the right or left margin of the table.
Pressing the >right< or >left< arrow keys once again after the input
cursor has reached the right or left table margin will move the cursor
one more step in the desired direction.
Press ESC to finish the measurement place name entry:
>Accept changes<
Saves the input
>Back<
Enables corrections
>Cancel<
Aborts the input and the previous name will
be kept
Fig. 29-4
Accepting the new measurement place name
>Diameter<
Enter the
accurate
inner diameter of the measurement place here.
Enter inside pipe diameter
The DN number is not equal to the inner pipe diameter which, depending on pressure level
and pipe material, may significantly vary from the DN number.
Entering an incorrect inside pipe diameter inevitably leads to a faulty calculation of the
cross-sectional area resulting in faulty calculated flow readings.
Maximum inside diameter depending on version
With NFP version “25” the maximum possible inside diameter is limited to 0.55 m. Version
“28” permits to specify diameters up to 0.85 m.
>Qmin<
This parameter is to suppress lowest evaluated and indicated vol-
umes which are either undesired or not relevant.
The parameter is mainly used in permanently full large pipelines with
partially low movements or to suppress indication and output of low
leakage volumes.
Q
min
: Measurement values lower than this one will be set to “0”.
Only positive values are allowed to be set. These values are going to
be considered as absolute values and therefore have positive as well
as negative effects.
>Operation mode<
This parameter is for adjustment to existing applications. Distinguish
between:
>Cyclic (pump)<
Medium condition changes between standing
and strong movement.
Primarily for use in applications such as
Содержание NFP
Страница 82: ...Instruction Manual NIVUS Full Pipe page 82 NFP rev 04 23 05 2018 Approvals and certificates...
Страница 83: ...Approvals and certificates NFP rev 04 23 05 2018 page 83...
Страница 84: ...Instruction Manual NIVUS Full Pipe page 84 NFP rev 04 23 05 2018...