Parameter Setting
NFP - rev. 04 / 23.05.2018
page 53
Fig. 29-11 Programming error output
29.5 Parameter Menu “Relay”
Fig. 29-12 Menu relay
This menu allows defining both functions as well as accompanying parameters (such as limit
values, duration of impulse and more) of individual relay outputs. Select parameter “relay
function” in order to indicate the available functions.
Toggle between relays using the >left< and >right< arrow keys.
Fig. 29-13 Relay sub menu
Possible relay functions:
>Function<
No relay function; default setting
>Flow limit relay<
Relay will energise if a flow limit value (to be set) has been exceeded
and will de-energise if flow falls below a second limit value (to be set).
>Relay mode<
It is possible to select between >normally
open< and >normally closed<. The relay is
going to energise if >normally open< has
been selected and the according value has
been reached, if >normally close< has been
selected the relay will energise immediately
after the parameter has been set and will de-
energise as soon as the according value has
been reached.