background image

 

 

Installation and Connection 

 

NFP - rev. 04 / 23.05.2018 

 

page 23 

 

 

 

radioactive radiation 

During installation works keep in mind that electronic components may be irreversibly dam-

aged due to electrostatic discharge. Therefore avoid intolerably high electrostatic charge dur-

ing installation by implementing appropriate grounding measures. 

Clear view door

 

The clear view door of the measurement transmitter is provided with a protection foil for pro-

tection during transport and from scratches during assembly. This protection foil has to be 

removed immediately after the assembly.  

 

UV radiation 

If the view door protection foil has been exposed to direct solar radiation for a long period, it 

cannot be removed easily. 

Should this be the case try to clean the door using spirit or car polish.  

If these measures should not be successful the clear view door can be replaced by NIVUS 

at extra costs. 

 

20.2  Enclosure Dimensions  

 

Fig. 20-1 

Wall Mount Enclosure  

 

 

Содержание NFP

Страница 1: ...F LOW p e r m an en t Instruction Manual Flow Measurement Transmitter NFP NIVUS Full Pipe Firmware Revision 4 00 Original Manual German 23 05 2018 Rev 04 23 05 2018...

Страница 2: ...fo nivus fr www nivus fr NIVUS Ltd United Kingdom Wedgewood Rugby Road Weston under Wetherley Royal Leamington Spa CV33 9BW Warwickshire Phone 44 0 8445 3328 83 nivusUK nivus com www nivus com NIVUS M...

Страница 3: ...Economic Area EEA this instruction manu al must be translated into the language of the country in which the device is to be used Should the translated text be unclear the original instruction manual...

Страница 4: ...1 Valuation of the accident level 9 2 2 Warning notices on the product option 10 3 Safeguards and Precautions 10 4 Liability disclaimer 11 5 Use in accordance with the requirements 11 6 User s Respons...

Страница 5: ...ser 34 22 General principles 34 23 Operator Panel 35 24 Display 35 25 Operation Basics 37 Parameter Setting 38 26 Basics of Parameter Setting 38 27 Operation Mode RUN 40 28 Display Menu EXTRA 43 29 Pa...

Страница 6: ...rmation on Simulation 66 31 3 CAL Menu Relays 67 31 4 CAL Menu Simulation 68 Parameter Tree Menus available 69 Troubleshooting 74 Maintenance and Cleaning 77 32 Maintenance 77 32 1 Maintenance interva...

Страница 7: ...t any time If you should have problems to understand information contained within this instruction man ual either contact the manufacturer or one of the distributors for further support The manu factu...

Страница 8: ...ation Text Parameter or Menu Indicates a parameter or a menu that is selected or described Reference to docu ment Refers to an accompanying documentation Table 1 Structural elements within the manual...

Страница 9: ...gree of risk Indicates a possible danger with medium risk which may result in a life threatening situa tion or severe bodily injury if it is not avoided CAUTION Warnings in low risk or property damage...

Страница 10: ...ards and Precautions Working with NIVUS instruments requires to observe and to follow the safety measures and precautions below generally and at any time These notes and warnings will not be repeated...

Страница 11: ...in Germany e g VDE reg ulations such as applicable Ex regulations as well as safety requirements and regulations in order to avoid accidents shall be observed All operations on the device which go bey...

Страница 12: ...zard Install the transmitter out of Ex zones The Ex approval of the sensors is part of the Technical Description and or Instructions Manual The approval is valid only in conjunction with the according...

Страница 13: ...agents and disposal Connections Operators shall make sure prior to operating the instrument that during installation and initial start up the local regulations such as regulations for electrical conne...

Страница 14: ...report shall be sent to NIVUS GmbH in Eppingen Incomplete deliveries shall be reported in writing either to your local representative or directly to the NIVUS head office in Eppingen within two weeks...

Страница 15: ...ress of the manufacturer CE label Information on type and series serial no if available Year of manufacture the first four digits of the serial number represent the year and the week number of manufac...

Страница 16: ...gory II Pollution degree 2 Altitude AC unit for use in altitudes up to 3000 m above MSL At relay voltages 150 V the use is restricted to an altitude of max 2000 m AC and DC units Operating temperature...

Страница 17: ...ace per channel Terminal no 53 54 Uo 9 9 V Io 130 3 mA Po 322 mW Linear characteristics Co 9 7 F Lo 0 15 mH Ui 10 1 V Ii 136 mA The maximum values also apply to concentrated capacitance inductors that...

Страница 18: ...type of device is indicated by the article number which can be found on a weath erproof label on the bottom of the enclosure From this article key the type of device can be specified NFP Type 25 Nomin...

Страница 19: ...city Measurement The piezo crystal has a slope to the flow direction and operates as a flow velocity sensor Here an ultrasonic burst with a defined angle is sent into the medium All the particles in t...

Страница 20: ...t tern results see Fig 18 3 At the same time slightly different reflections will emerge addi tionally resulting from the fact that some particles have been turning around and thus have another shape o...

Страница 21: ...indicated on the display of NFP Fig 18 4 Investigated flow profile If there are appropriate banking distances on the measurement place available based on the known pipe geometry and the velocity dist...

Страница 22: ...resistance with 500 kOhm min will be required The cross sectional dimension of the power supply wires must be 0 75 mm 2 0 03 in 2 and must be in accordance to IEC 227 or IEC 245 The device protection...

Страница 23: ...the measurement transmitter is provided with a protection foil for pro tection during transport and from scratches during assembly This protection foil has to be removed immediately after the assembly...

Страница 24: ...tal pipe Avoid unnecessary movements to reduce the risk of building up static electricity Transport statically sensitive components in antistatic containers or packing materials To discharge your body...

Страница 25: ...tallation 20 5 1 Transmitter Connection Important Note The complete measurement system shall be installed and put into operation by qualified personnel exclusively In terms of electric connection plea...

Страница 26: ...lied lid and both screws respectively When connecting the transmitter observe the terminal clamp specifications listed below Power supply terminal 1 5 Spring cage terminal wire up to 2 5 mm strand up...

Страница 27: ...d units permit to tap the 24 V auxiliary voltage which is also available on 24 Volt DC 28 and 0 V DC 39 from connections DC DC clamps 4 and 5 too AC units therefore have a low impedance connection bet...

Страница 28: ...onnections Improper connections which lead to higher transition resistance or the use of other cables may lead to disturbance and errors in the measurement Disregarding may lead to errors or measureme...

Страница 29: ...tch position in terminal clamp housing Operation with AC voltage DC voltage A DC unit can be operated exclusively with 24 V 15 DC voltage An AC unit can be operated exclusively with 100 240 V 10 15 AC...

Страница 30: ...resistance must be taken into account consider ing the allowed total resistance Observe the Technical Instructions for Correlation Sen sors and external Electronic Box Overvoltage protection elements...

Страница 31: ...24 V and DataPro 2x1 24 V 24 V Tr Permissible cable length If using the sensors in Ex areas consider the connected loads of the overvoltage protection devices as well as capacity and inductance of the...

Страница 32: ...nual NIVUS Full Pipe page 32 NFP rev 04 23 05 2018 Fig 20 9 Connecting the overvoltage protection for power supply and analog out puts Fig 20 10 Connecting the overvoltage protection for velocity sens...

Страница 33: ...Installation and Connection NFP rev 04 23 05 2018 page 33 Fig 20 11 Connecting the overvoltage protection for velocity sensor Non Ex version...

Страница 34: ...nsmitter Connection In case of doubt regarding installation connection or the setting of parameters contact our hotline 49 0 7262 9191 955 To put the entire system into operation additionally consult...

Страница 35: ...nfirmation button Fig 23 1 Operator panel 24 Display The NFP has a large back lit graphic display with a resolution of 128x64 pixel This ensures a comfortable communication mode for the user 1 Name of...

Страница 36: ...ost extensive menu for initial start up Parameter setting of dimensions of the measurement Parameter setting of sensors Parameter setting of analog and digital outputs Damping settings System reset I...

Страница 37: ...name Select preset values e g units m cm l s m s etc Decrease values Set decimal point left Press once to toggle between indication mode and overview menu main menu Jump across in main or submenu Jum...

Страница 38: ...t The system PIN protects from unauthorised access Single input of system PIN within 24 hours The system PIN must be entered only once within 24 hours The transmitter will save more password protected...

Страница 39: ...of the NFP Depending on the device type however there might be only one analog output available or the maximum pipe diameter to set might be 450 mm see Table 3 After mounting and installing sensor an...

Страница 40: ...last key action 27 Operation Mode RUN This menu is a display menu for standard operation mode and indicates day values and error messages Containing the following sub menus it is not required for para...

Страница 41: ...2 Time of day totalising cycle Information concerning the totalising If the transmitter is disconnected from mains at the time of totalising set it is not possible to create or to save a total for th...

Страница 42: ...ective Sensor status sensor reports status error Temperature medium temperature significantly going over or falling below a range of 10 C of the permissible temperature range 40 100 C or defective tem...

Страница 43: ...down keys Info This point provides comprehensive information on unit type transmit ter serial no and software version see Fig 28 2 The menu is sub divided into 4 single pages Pressing the key right an...

Страница 44: ...he display format Language Select from German English French and Polish Display Permits to adjust the display contrast Use the arrow key down to decrease and up to increase the value in steps of 5 The...

Страница 45: ...is to show the same total value Specify new total value Confirm with Enter key Enter system PIN and confirm New total value is shown in the main screen Fig 28 6 Modification of totalizer Modify PIN Sy...

Страница 46: ...hence restores transmitter access Fig 28 7 Submenu modify PIN 29 Parameter Menu PAR This menu is used to set all relevant parameters required to ensure reliable device function Choose from the followi...

Страница 47: ...Underneath the measurement place name you can find a table with 20 lines containing all uppercase and lowercase letters as well as a large number of special characters see Fig 29 3 Use the keys up and...

Страница 48: ...Entering an incorrect inside pipe diameter inevitably leads to a faulty calculation of the cross sectional area resulting in faulty calculated flow readings Maximum inside diameter depending on versi...

Страница 49: ...upstream unlike heading downstream to wards the flow direction as in standard applications but is to detect positive velocities however This is the only case which requires set ting negative here 29...

Страница 50: ...y critical applications it is pos sible to externally switch over between oper ation modes cyclic pump cont opera tion via the digital input Logic Toggle between inverse and non inverse mode Default n...

Страница 51: ...put of the measured average velocity Temperature Output of the measured medium tempera ture Constant current The analog output can be set to output a constant current which is independent from any rea...

Страница 52: ...log signal to decrease positive flow will cause the sig nal to increase Value at 20mA Enter the measurement value at 20 mA Error mode Options active not active If this mode has been unlocked by settin...

Страница 53: ...eys Fig 29 13 Relay sub menu Possible relay functions Function No relay function default setting Flow limit relay Relay will energise if a flow limit value to be set has been exceeded and will de ener...

Страница 54: ...ired and the limit value is present yet If the value falls below the limit threshold for a moment the cycle will begin anew Name The relay output name may consist of four characters max which will be...

Страница 55: ...olume will be added internally until the value set has been reached Then an impulse sig nal with the duration set will be output and the internal counter will be set to 0 again The course of events wi...

Страница 56: ...t yet If the value falls below the limit threshold for a moment the cycle will begin anew OFF delay The OFF event in case of reaching the limit value can be delayed by up to 9999 seconds max The relay...

Страница 57: ...execute a gen eral reset The unit will be in initialising mode subsequently which requires the operation language to be set The transmitter will now overwrite the flash memory restarting the program D...

Страница 58: ...5 10 15 20 30 minutes or 1 hour Default setting 1 minute Format of num bers Use of comma or dot XX YYY or XX YYY for data storage Default setting XX YYY Unit system Define the units unit system used...

Страница 59: ...The menu therefore primarily serves in order to assess the parameter settings and for error diag nosis Indication of all theoretically possible inputs and outputs The menu basically allows indicating...

Страница 60: ...nverter for output is indicated here This menu cannot be used to detect and to view external incorrect connections 30 3 I O sub menu Relay This submenu indicates the conditions calculated by the trans...

Страница 61: ...aximum permissible memory 32 GB Working with USB stick Plug the USB stick into the USB slot located next to the keypad Functions Transmission of measurement data to USB stick Instrument parameters can...

Страница 62: ...r is transmitted to USB stick File name PAR_DATE_TIME csv Changed only Only modified other than the default set tings parameters are transmitted to USB stick File name CHGPARAM_DATE_TIME csv Load para...

Страница 63: ...mperature Fig 30 7 v Sensor Use left and right arrow keys to browse through more screens The velocity graph represents the velocity distribution of individual gates within the pipeline This graph enab...

Страница 64: ...max should not exceed 25 40 dB significantly Fig 30 11 Sensor noise In the event of higher values check the layout of the sensor cable as well as the transmitter earthing 30 8 I O sub menu v Gate Tabl...

Страница 65: ...re the negative flow direction max velocity Defines the maximum flow velocity range measured and evaluated by the transmitter Default setting 6 m s Positive and negative velocities are not measured If...

Страница 66: ...l injury or property damage The simulation of NFP outputs will access any following facility areas without any safety locking measures Simulations may only be carried out by NIVUS expert personnel or...

Страница 67: ...e Fig 31 4 2 Fig 31 4 Simulation of analog outputs 31 3 CAL Menu Relays Relays Selecting the Relay Outputs option requires to enter the system PIN once again to make sure that simulations can be carri...

Страница 68: ...flow value which has been calculated by means of the simulated readings will be indicated on the bottom line of the screen Relays which might have been set will switch and programmed mA outputs suppl...

Страница 69: ...e Erase counter Error message Parameter menu PAR Default settings PAR Measure ment Place Name NIVUS Diameter 0 45 Qmin 0 Operation mode Cyclic pump x Cont opera tion Velocity Install Direc tion Positi...

Страница 70: ...delay 0 Name R1 Vel limit relay Relay mode Normally open x Normally closed ON point OFF point ON delay 0 OFF delay 0 Name R1 Temp limit relay Relay mode Normally open x Normally closed ON point OFF po...

Страница 71: ...point ON delay 0 OFF delay 0 Name R1 Impulse duration 5 Volume impulse 1 Error message Relay mode Normally open X Normally closed ON point OFF point ON delay 0 OFF delay 0 Name R1 Setup pa rameter Sys...

Страница 72: ...nalog Out puts Relays Data storage USB Info Erase USB Stick Store NivuSoft Store CSV Save parameter Load parameter Measure data v Sensor v Sensor noise v Gate Calibration menu CAL Default settings CAL...

Страница 73: ...o 3 Info 4 Unit system Metric x UK english US english Units Flow rate Veloc Total Display format Flow rate Veloc Total Language Deutsch x English Francais Polski Display Contrast 50 System time Info S...

Страница 74: ...ive flow direction Check sensor installation direction rotate sensor if required If measurement fails only with the flow direction reverted go to CAL Flow Velocity min max value set min value to 6 0 m...

Страница 75: ...f cables are crushed for extensions cable types short circuits surge arresters or improper resistive loads Check echo profile flow velocity signal cable pa rameters and temperature in I O menu Check i...

Страница 76: ...nd connections overvoltage pro tection elements as well as input and output clamps Check input range 0 or 4 20 mA of following system Check input span of following system Check possible offset of foll...

Страница 77: ...strial Safety Regulations are unaffected NIVUS recommends having the entire measurement system inspected by the NIVUS cus tomer service once per year Depending on the area of use the maintenance inter...

Страница 78: ...hese hints can be found in the Technical Instruction and or the Instructor Manual This these document s is are part of the standard sensor delivery 34 Dismantling Disposal Improper disposal may be har...

Страница 79: ...or in the valid price list 36 Accessories Mounting stand ZMS0 151 Mounting stand 1 material 1 4301 V2A height 1700 mm incl base plate and fastening tube Assembly plate ZMS0 161 Assembly plate for moun...

Страница 80: ...eption 22 Info 41 ESD 24 Ex approval 16 Ex protection 12 F Flow Velocity Measurement 19 Front panel 24 26 Functional Principle 19 G General Reset 57 Germ contamination 10 Graphic Display 15 I Improper...

Страница 81: ...66 Spare parts 79 Specifications 16 Start screen 39 Storage temperature 16 Storing 14 Symbols 9 T Terminal Clamp Housing 15 Terminal housing 24 26 Translation 3 Transmitter Connection 25 Transmitter I...

Страница 82: ...Instruction Manual NIVUS Full Pipe page 82 NFP rev 04 23 05 2018 Approvals and certificates...

Страница 83: ...Approvals and certificates NFP rev 04 23 05 2018 page 83...

Страница 84: ...Instruction Manual NIVUS Full Pipe page 84 NFP rev 04 23 05 2018...

Страница 85: ...Approvals and certificates NFP rev 04 23 05 2018 page 85...

Страница 86: ...Instruction Manual NIVUS Full Pipe page 86 NFP rev 04 23 05 2018...

Страница 87: ...Approvals and certificates NFP rev 04 23 05 2018 page 87...

Отзывы: