-8-
Key operation
Display
Turn the power off
PEAK
ZERO
MEM
UNIT
1. Hold down PEAK and UNIT key,
2. Press POWER key.
PEAK
ZERO
MEM
UNIT
Release POWER key only.
4.5.Tracking
A load cell strain gauge is used for the FGV force sensing. This type of sensor can alter its output due to changes in temperature.
The tracking feature algorithm can negate this slight change. When measuring very small forces, an error could arise while tracking
is on. It is recommended on light load sensing to turn the tracking function off. To turn the tracking off, hold the PEAK key and UNIT
key simultaneously while the power is off. Next, press and release the POWER key. Then, release the PEAK key and UNIT key after a
value has remained on the screen for 1 second.
Display force gauge model
Standard measuring mode
Please release the
PEAK key and UNIT key
after the value appears
in the display more than
1 second.
Tracking ON
Tracking OFF
Tracking OFF
Tracking ON
Turn power off, then press POWER key, you can confirm the condition
of the present tracking.
1.
2.
FGV
Key operation
Display
Turn the power off
PEAK
ZERO
MEM
UNIT
1. Hold down UNIT key
2. Press POWER key.
PEAK
ZERO
MEM
UNIT
Release POWER key only.
In the case you attach FGV with a stand, display of the value and the units can be reversed in order to read the display easily.
Turn POWER off. Press the UNIT key and hold, then press POWER key and release (release UNIT key after the value appears in the
display more than 1 second). Then you can reverse the disply.
4.6.Reverse the display
Please release the
UNIT key after the
value appears in the
display more than 1
second.
Display force gauge model
Standard display
Standard measuring mode
Reverse display
Reverse disply
Standard display