-11-
Key operation
Display
PEAK
ZERO
MEM
UNIT
Press UNIT to cycle
through settings
ZERO
Save current setting and return to the standard measuring mode.
PEAK
Save the change and then move to f06.
5.5.Response Time: f05
The response time function smooths out the gauge's sampling, and adjustis the sampling period accordingly
The available response times are 3, 20, and 150 msec.
Press UNIT to change the setting, PEAK to save and move to the next function, or ZERO to save and finish.
Response time 3msec Response time 20msec Response time 150msec
* Filter response time show 90% of step input.
Depend on the filter response, sampling period
and analog output update period is decided.
Input (change of actual load)
Measuring value
90%response time
90%
Filter response
Sampling period• Analog output update
period
3msec
1msec
20msec
1msec
150msec
6.7msec
Key operation
Display
PEAK
ZERO
MEM
UNIT
Press UNIT to cycle
through settings
ZERO
Save current setting and return to the standard measuring mode.
PEAK
Save the change and then move to f07.
5.6.External output : f06
It's available to change the external output; RS-232C, USB and Digimatic output.
Press UNIT key to switch the external output. Press PEAK key to save and move to next function.
RS-232C • USB output
Digimatic output