-20-
Key operation
Display
Turn power off
PEAK
ZERO
MEM
UNIT
1. Hold down MEM key
2. Press POWER key
PEAK
ZERO
MEM
UNIT
Release POWER key only
MEM
Return to memory data display.
PEAK
Output the memory data with RS-232C, and then move to standard measuring mode.
Finish display of the memory data. (Turn power off)
6.7.1.2.Statistics memory data
1. When in memory mode, UNIT will cycle through the available statistics data.
2. Each press of UNIT will switch between the following items: positive maximum value, negative maximum value, positive minimum
value, negative minimum value, positive peak value negative peak value, average value, standard deviation.
3. Press MEM to exit to the recorded measurement values.
4. Press PEAK key to output the recorded memory via RS-232C. Refer to section 7 regarding RS232C communication.
Plus Maximum value
Minus Maximum value
Plus Minimum value
Standard deviation
(Show 2 digits after the decimal point)
UNIT
UNIT
UNIT
UNIT
UNIT
UNIT
UNIT
UNIT
Minus Minimum value
Plus Peak value
Minus Peak value
Average
Memory number
Display one after the other
Memory data
UNIT
Display statistical data
Please release the
MEM key after the
value appears in the
display more than 1
second.
1.
2.
MEM
Содержание FGV-HXY
Страница 34: ...33...