FGV-HXY Digital Force Gauge

Operation Manual

High Capacity Digital Force Gauges with 180° reversible dis-
play & dual labeled keypad for both push, pull applications.

 

The Series FGV-HXY Digital Force Gauges offer an enhanced 
heavy-duty housing allowing them to be utilized in high capac-
ity tension and compression applications. These instruments 
are utilized in a variety of applications such as, incoming quality 
inspection, finished goods testing, R&D or almost any portable 
or force stand testing requirement. 

Models are available in 500 lb or 1000 lb ranges providing nec-
essary performance for high capacity testing.   

The 180° reversible display combined with the adjustable force 
direction and dual labeled keypad enable these units to excel 
in both compression and tension applications in any direction.
FGV models possess onboard statistical calculations and pro-
vide fast access to maximum, minimum, average and standard 
deviation data. Models include the free Toriemon software and 
USB cable allowing further data analyis on a PC. The FGV-
HXY’s 1,000 point memory allows for data exporting in real time 
or after testing is complete. In addition, the RS-232C, USB and 
analog outputs provide maximum communication flexibility.

Do not operate or store instrument in the following 

locations:

Explosive areas, near water, oil, dust or chemicals; ar-

eas where the temperatures can exceed 104°F 

(40° C)

Take precaution to not drop the force gauge.  Damage 

to the sensor may occur.

Do not modify, disassemble or attempt to repair the 

unit.  Send to the factory for proper repair.

If upon delivery damage to the unit is detected, do not 

operate the unit.  Notify the shipping carrier immedi-

ately to obtain damage claim instructions.

Only measure forces that are in line with the measur-

ing shaft.  Do not attempt to take any measurements 

(tension or compression) at any angle.  Failure to keep 

measurements in line will damage the instrument. 

Capacity: 

-500HXY: 500 lb, 250 kg, 2500 N; -1000HXY: 1000 

lb, 500 kg, 4448 N

Resolution: 

0.1 lb/0.1 kg/1 N

Accuracy: 

±0.2% F.S.

Display: 

Four digit LCD, .47” high (12 mm) with various indi-

cators including tension and low battery indication 
(reversible) Minus sign for tension

Average/Peak Mode:

 Selectable

Display Update: 

0.05, 0.1, 0.2, 0.3, 0.5, 1 sec

Sampling Rate: 

1000 Hz

Overload Capacity: 

150% of F.S.

Power: 

Rechargeable Ni-Cad batteries (included) or AC 

adapter/charger (included)

Auto Power Shut-Off: 

Yes (not active if adapter/charge is 

in use)

Battery Charge: 

17 hours (when fully charged)

Recharge Time: 

Approximately 18 hours (when fully dis-

charged)

Temperature Range: 

32 - 104°F (0 - 40°C)

Dimensions: 

8.43” x 3.23” x 2.25”  (214 x 82 x 57 mm)

Product Weight: 

Approx. 2 lb (900 g) for 500 lb range; 2.65 

lb (1.20 kg)

 

for 1000 lb range

Package Weight: 

Approx. 7.45 lb (3.38 kg) for 500 lb range;  

8.25 lb (3.74 kg) for 1000 lb range

Certification: 

CE, RoHS

Warranty: 

2-Year

Included Accessories: 

AC adapter/charger, carrying case, 

2 handles, metric thread to inch adapter and attachments (flat 
head, hook, chisel, notched head, cone head, extension rod), 
USB cable. Note: software available at www.shimpoinst.com 
for free download

Содержание FGV-HXY

Страница 1: ...e to the sensor may occur Do not modify disassemble or attempt to repair the unit Send to the factory for proper repair If upon delivery damage to the unit is detected do not operate the unit Notify t...

Страница 2: ...ng Caution Warning Caution Safety Precaution Serious injury or possibly even death may result if the gauge is used wrongly A serious problem or issue may result if the gauge is used wrongly Take preca...

Страница 3: ...ct load or excessive axial forces Bending direction Twisting direction Accuracy performance degrades overtime It is recommended the force gauge be checked for performance regularly The frequency of th...

Страница 4: ...7 6 6 2 Storing data 18 6 6 2 1 Continuous memory mode 18 6 6 2 2 Single memory mode 18 6 6 2 3 Standard memory mode 18 6 7 Recalling memory data 19 6 7 1 Continuous memory mode 19 6 7 1 1 Measuring m...

Страница 5: ...ith USB 6 9 USB communication 1000 data point memory 6 6 Memory Comparator shows pass or fail I O output of the result 6 5 Comparator Reverse the display of the measuring value and the unit 4 6 Revers...

Страница 6: ...ng 7 POWER key Press the POWER and release to turn ON OFF It is used to reverse the display Tracking ON OFF It is used for comparator memory setting 8 MEM key Press the MEM key under the standard meas...

Страница 7: ...of battery BAT is shown Even if turn the power off during the charge BAT will be displayed PEAK turns on when plus peak hold mode PEAK turns on when m i n u s p e a k h o l d mode Newton Depend on the...

Страница 8: ...ours per 1 full charge 2 The nickel hydride battery is charged automatically when it is discharged during the usage of AC adapter If you charge the battery frequently its lifetime will be shortened Wh...

Страница 9: ...Standard measuring mode Please release the PEAK key and UNIT key after the value appears in the display more than 1 second Tracking ON Tracking OFF Tracking OFF Tracking ON Turn power off then press...

Страница 10: ...s 0001 Display update time f02 1 2 3 5 10 20 times second 3 Auto power off f03 10 10 minutes oFF not valid 10 RS 232C baud rate f04 2400 4800 9600 19200 bps 2400 Response Time f05 3 20 150 msec 3 Exte...

Страница 11: ...easuring mode PEAK Save the change and then move to f04 5 3 Auto power off f03 If the gauge is on and there is no activity for 10 minutes the unit automatically powers off to conserve battery charge W...

Страница 12: ...e 150msec Filter response time show 90 of step input Depend on the filter response sampling period and analog output update period is decided Input change of actual load Measuring value 90 response ti...

Страница 13: ...key and UNIT key after the value appears in the display more than 1 second ZERO POWER Function mode Turn POWER off Press ZERO key and hold then press POWER key and release release ZERO key after the v...

Страница 14: ...nder the plus peak hold mode or minus peak hold mode The plus peak value or minus peak value is cleared In plus peak hold or minus peak hold modes the tare cannot be performed even when pressing and r...

Страница 15: ...ry mode setting MEM Set the memory mode single mode continuous mode standard mode Std When you set 0 at both HI limit and LOW limit comparater function does not work 6 5 3 Setting HI limit 1 Press the...

Страница 16: ...n output signal of comparator LO limit 6 6 Memory There are 3 modes at memory mode as follows Continuous memory Allows the recording of up to 1000 data points The recording starts when you push MEM an...

Страница 17: ...N MIN Continuous memory mode Ex1 Ex2 Ex3 MAX MAX MIN 0 MIN 0 MAX MAX MAX MIN Start memory Start memory Start memory Finish memory Finish memory Finish memory 0 0 0 MIN 0 MAX 0 Average value AVE Averag...

Страница 18: ...it 0 Comparator LO limit setting LO Set comparator LO limit 0 Memory mode setting MEM Set memory mode Single mode Continuous mode Standard mode Std When you set 0 at both HI limit and LO limit compara...

Страница 19: ...umber arrives at 1000 6 6 2 2 Single memory mode 1 During the standard measuring mode press MEM key M turns on at the unit display and one data is recorded 2 If 100 data points are recorded FULL appea...

Страница 20: ...orded measurement value of that block 2 Press MEM to review the previous data block recorded 3 Press UNIT to cycle through the available recorded statistics In Continuous mode the available statistics...

Страница 21: ...egative maximum value positive minimum value negative minimum value positive peak value negative peak value average value standard deviation 3 Press MEM to exit to the recorded measurement values 4 Pr...

Страница 22: ...s and release POWER Release MEM when you see Slg on the display The unit will alternate between showing the data block number and the recorded measurement value of that block 2 Press MEM to review the...

Страница 23: ...ing RS232C communication Key operation Display Turn power off PEAK ZERO MEM UNIT 1 Hold down MEM key 2 Press POWER key PEAK ZERO MEM UNIT Release POWER key only MEM Return to memory data display PEAK...

Страница 24: ...a greater detail in the next section 4 Press PEAK to output the recorded memory via RS 232C Refer to section 7 regarding RS232C communications Key operation Display Turn power off PEAK ZERO MEM UNIT...

Страница 25: ...itive minimum value negative minimum value positive peak value and negative peak value 3 Press MEM to exit to the recorded measurement values 4 Press PEAK key to output the recorded memory via RS 232C...

Страница 26: ...USB 6 9 2 Download ToriemonUSB If you connect the force gauge to a PC with the USB cable you can download data to the PC from the force gauge of the current values being measured or from data that is...

Страница 27: ...put can be set by external output setting f06 of function mode 7 2 RS 232C Output You can operate this equipment from your PC if connecting it by using optional cable for RS 232C Cannot use USB commun...

Страница 28: ...t and 4 digit number BBcr Request for continuous transmission of measuring data 10 times second BBcr NAoooooocr BB1cr Request for continuous transmission of measuring data 20 times second BB1cr NAoooo...

Страница 29: ...rnal output setting f06 function mode Maximum allowance voltage DC 30V current 5mA Please connect the power and load to avoid going over maximum allowance Overload output When the overload condition i...

Страница 30: ...limit output 8 ZERO signal input Input signal 9 HOLD signal input 10 PEAK signal input 11 Common for comparator output 12 GND The device printer provided for Mitutoyo Digimatic Please purchase the opt...

Страница 31: ...gnal input 9pin ZERO signal input 10pin GND 12pin Signal is ON DC 3 3V 0 5mA 2 HOLD signal input HOLD signal is ON when HOLD signal input pin 9 pin and GND 12 pin is changed from OPEN to CLOSE The dis...

Страница 32: ...it Questions Explanation Reference How long does rechargeable battery Nickel hydride battery last Enable to use 500 times or more by complete electric discharge It depends on the status of use Please...

Страница 33: ...CHECKED UNITS SCALE TITLE Toshi MO YS mm 1 1 MODEL ORIGINAL DRAWING NO DRAWING NO BGBM068 FGC H FGV H FGN H FGE H High Load Force Gauge Attachment Drawing M10 EJ350280 DATE 04 24 2017 Itasca IL 60143...

Страница 34: ...33...

Отзывы: