![NI PXI-4220 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/ni/pxi-4220/pxi-4220_user-manual_1685516046.webp)
Chapter 4
Theory of Operation
©
National Instruments Corporation
4-11
NI PXI-4220 User Manual
•
Sensitive to bending strain.
•
Rejects axial strain.
•
Compensates for specimen temperature variation.
•
Sensitivity at 1000
µε
is
∼
1 mV
out
/
V
EX
input.
Figure 4-9.
Half-Bridge Type II Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•
R
1
and R
2
are half-bridge completion resistors.
•
R
3
is the active strain-gauge element measuring compressive
strain (–
ε)
.
•
R
4
is the active strain-gauge element measuring tensile strain (+
ε
).
•
V
EX
is the excitation voltage.
•
R
L
is the lead resistance.
•
V
CH
is the voltage measured.
To convert voltage readings to strain units, use the following equation:
where
•
R
g
is the nominal gauge resistance.
•
R
L
is the lead resistance.
•
GF
is the gauge factor.
To simulate the effect on strain of applying a shunt resistor across R
3
,
use the following equation:
R
1
R
2
+
–
–
+
V
EX
V
CH
R
4
(+ )
R
3
(– )
R
L
R
L
R
L
strain
ε
( )
2
–
V
r
GF
------------
1
R
L
R
g
------
+
×
=
ε
s
2
U
–
GF
----------
=
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com