Chapter 2
Connecting Signals
NI PXI-4220 User Manual
2-12
ni.com
Remote Sense
For remote sensing, wire the NI PXI-4220 for remote sense as shown in
Figure 2-11. Refer to the
Remote Sense
section of Chapter 4,
Theory of
Operation
, for more information about remote sensing. Using remote sense
with the NI PXI-4220 is recommended to properly regulate the excitation
voltage,
V
EX
, being applied to the sensor.
Figure 2-11.
Remote Sense Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•
R
1
, R
2
, R
3
, and R
4
are the transducer bridge elements.
•
V
CH
is the measured voltage.
•
V
EX
is the excitation voltage.
•
R
L
is the lead resistance.
•
V
RSENSE
is the measured bridge excitation.
Note
Remote sense is illustrated using a full-bridge circuit diagram for simplicity, but in
general, the
V
RSENSE
terminals should be connected to your transducer at the points that
supply
V
EX
. For example, connect RS
X
+ (pin 3) to P
X
+ (pin 2), and RS
X
– (pin 8) to
P
X
– (pin 7) at the sensor.
+
–
+
–
R
L
R
L
NI PXI-4220
Transducer
R
1
R
2
R
4
R
3
S+ (Pin 1)
S– (Pin 6)
P+ (Pin 2)
P– (Pin 7)
RS+ (Pin 3)
RS– (Pin 8)
V
EX
V
RSENSE
+
–
Remote
Sense
Feedback
Loop
V
CH