![NI PXI-4220 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/ni/pxi-4220/pxi-4220_user-manual_1685516018.webp)
Chapter 2
Connecting Signals
NI PXI-4220 User Manual
2-4
ni.com
Figure 2-3 shows the location of the socketed NI PXI-4220 resistors.
Figure 2-3.
Socketed Resistor Locations
Quarter-Bridge Type I
This section provides information about connecting the quarter-bridge
strain-gauge configuration type I.
Figure 2-4 shows the quarter-bridge type I circuit wiring diagram. Refer to
the
Quarter-Bridge Type I
section of Chapter 4,
Theory of Operation
, for
more information.
Note
If you have a two-wire strain-gauge element, connect the active element leads to
S+ (pin 1) and P+ (pin 2) and short S+ (pin 1) to QTR/SCB (pin 9) and SCA (pin 4).
1
R4 (Shunt Cal B CH 0)
2
R1 (Shunt Cal A CH 0)
3
R5 (Quarter-Bridge Completion CH0)
4
R2 (Quarter-Bridge Completion CH1)
5
R6 (Shunt Cal B CH 1)
6
R3 (Shunt Cal A CH 1)
PXI-9-PIN DSUB