Appendix D
Common Questions
NI PXI-4220 User Manual
D-2
ni.com
increase offset error beyond the limits given in the specifications. You
cannot perform offset null compensation when the excitation is set to 0 V.
How do I program the NI PXI-4220?
Refer to Chapter 5,
Developing Your Application
, for application
programming information or your ADE help file. There is no register-level
programming manual available for the NI PXI-4220.
If I am powering my bridge-based transducers with an external
voltage source, what voltage setting should I set on the NI PXI-4220?
Even with an external voltage source for the transducers, internal excitation
voltage is still used by the offset null compensation circuitry inside the
NI PXI-4220. Set the internal excitation voltage equal to the external
excitation voltage to ensure the accuracy of offset null compensation.
How do I perform an external calibration of the NI PXI-4220?
As of the NI PXI-4220 release, an external calibration document is not
available. To see if an NI PXI-4220 external calibration document is
currently available, click
Manual Calibration Procedures
at
ni.com/calibration
.