47
New Brunswick Scientific
User’s Guide
Figure 28: DO Probe in Adapter Plug (cross-section)
7.3.4
DO Probe Polarization
#
NOTES:
If the probe has been disconnected from a voltage source (either the unit’s
O
2
amplifier or a separate polarizing module) for longer than 5 minutes, it
will need to be re-polarized.
To re-polarize:
Connect the probe to the operating O
2
amplifier (or polarizing module).
Allow SIX HOURS FOR POLARIZATION prior to calibrating the probe.
7.3.5
DO Probe Calibration: Setting Zero
#
NOTE:
The DO probe is calibrated after sterilization.
There are two methods to obtain zero for calibrating the DO probe. Review
both methods and use the one you prefer:
Method 1:
1.
Remove the DO cable from the DO electrode.
O-RING
Содержание BioFlo 415
Страница 12: ......
Страница 128: ...116 BioFlo 415 y M1360 0050 User s Guide 2 20 0 D DR RA AW WI IN NG GS S Figure 65 Controller Schematics overview ...
Страница 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Страница 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Страница 145: ...133 New Brunswick Scientific User s Guide ...
Страница 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Страница 147: ...135 New Brunswick Scientific User s Guide ...
Страница 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Страница 149: ...137 New Brunswick Scientific User s Guide ...
Страница 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Страница 151: ...139 New Brunswick Scientific User s Guide ...
Страница 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Страница 153: ...141 New Brunswick Scientific User s Guide ...
Страница 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Страница 155: ...143 New Brunswick Scientific User s Guide ...
Страница 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Страница 157: ...145 New Brunswick Scientific User s Guide ...
Страница 158: ...146 BioFlo 415 y M1360 0050 User s Guide ...
Страница 164: ...152 BioFlo 415 y M1360 0050 User s Guide ...