3.4 Temperature
Control
The culture temperature setpoint may be selected within the range from 5
°
C above
coolant temperature to 80
°
C (
±
0.1
°
C). It is controlled by the process-control
software. The media temperature is sensed by a Resistance Temperature Detector
(RTD) submerged in the thermowell.
Default P & I (proportional & integral) values are preset at the factory.
We strongly
recommend that you maintain the factory-set parameters.
3.5 Aeration
Up to four gases, including air, nitrogen, carbon dioxide and oxygen, can be
introduced into the media through the ring sparger. The flow rate is controlled
automatically by one, two, three or four thermal mass flow controller(s), according to
the definition of your system. The thermal mass flow controller is regulated
automatically according to values set via the control cabinet touchscreen. If you wish
to use a rotameter, there is an option for no thermal mass flow controller.
The percentage of oxygen blended with the sparge air can be controlled manually by
the user or automatically through the controller by applying the O
2
enrichment
function.
(For further information on cascading, see Section 11.)
3.6 pH
Control
pH is controlled in the range of 2.00-12.00 (
±
0.01). The pH is sensed by a Gel-filled
pH probe
(see Figures 32a & 32b)
. Control is maintained by a P & I (proportional &
integral) controller which operates two peristaltic pumps, assigned to acid and base
addition ports, or controls the use of gas(es) for this purpose. The user can also select
a deadband value to control pH within the user-assigned range: no acid or base will be
added when the pH value falls within the deadband tolerance above or below the
setpoint.
Default P & I (proportional & integral) values are preset at the factory.
We strongly
recommend that you maintain the factory-set parameters.
Содержание BioFlo 415
Страница 12: ......
Страница 128: ...116 BioFlo 415 y M1360 0050 User s Guide 2 20 0 D DR RA AW WI IN NG GS S Figure 65 Controller Schematics overview ...
Страница 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Страница 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Страница 145: ...133 New Brunswick Scientific User s Guide ...
Страница 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Страница 147: ...135 New Brunswick Scientific User s Guide ...
Страница 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Страница 149: ...137 New Brunswick Scientific User s Guide ...
Страница 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Страница 151: ...139 New Brunswick Scientific User s Guide ...
Страница 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Страница 153: ...141 New Brunswick Scientific User s Guide ...
Страница 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Страница 155: ...143 New Brunswick Scientific User s Guide ...
Страница 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Страница 157: ...145 New Brunswick Scientific User s Guide ...
Страница 158: ...146 BioFlo 415 y M1360 0050 User s Guide ...
Страница 164: ...152 BioFlo 415 y M1360 0050 User s Guide ...