54
BioFlo 415
y
M1360-0050
User’s Guide
Figure 22: Sterilization Screen
6.
Press the
Start
button to initiate the sterilization sequence.
7.
The system will automatically open and close the valves on the unit to
ensure that the system is sterilized properly.
#
NOTE:
The air inlet and exhaust filter are connected to a solenoid valve, by
means of a cable that should be connected to the controller. This valve
pulses during sterilization mode to ensure sterilization of the inlet and
exhaust filter.
#
NOTE:
The sample valve should be sterilized separately in an autoclave, then
aseptically connected to the unit in a clean environment.
8.
When the sterilization hold phase (“Steril. Time”) ends, the operator should
reopen the needle valve on the water return line from the exhaust line.
WARNING!
All metal surfaces remain very hot after sterilization. Be sure to wear
protective gloves when you move the sparge valve from
sterilization to sparge position.
Содержание BioFlo 415
Страница 12: ......
Страница 128: ...116 BioFlo 415 y M1360 0050 User s Guide 2 20 0 D DR RA AW WI IN NG GS S Figure 65 Controller Schematics overview ...
Страница 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Страница 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Страница 145: ...133 New Brunswick Scientific User s Guide ...
Страница 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Страница 147: ...135 New Brunswick Scientific User s Guide ...
Страница 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Страница 149: ...137 New Brunswick Scientific User s Guide ...
Страница 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Страница 151: ...139 New Brunswick Scientific User s Guide ...
Страница 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Страница 153: ...141 New Brunswick Scientific User s Guide ...
Страница 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Страница 155: ...143 New Brunswick Scientific User s Guide ...
Страница 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Страница 157: ...145 New Brunswick Scientific User s Guide ...
Страница 158: ...146 BioFlo 415 y M1360 0050 User s Guide ...
Страница 164: ...152 BioFlo 415 y M1360 0050 User s Guide ...