111
New Brunswick Scientific
User’s Guide
18.3
Vessel & Tubing
After each and every run, clean the vessel and the headplate with its associated parts.
All tubing should be inspected and replaced as needed.
18.4 Periodic
Inspection
At three-month intervals
, perform the following checks and inspections.
#
NOTE:
Before you begin, make sure that the power switch is in the OFF position
and that the power supply has been disconnected.
1.
Check all controls and accessible items (power switch, connectors, screws, nuts
and bolts) to make sure they are properly tightened. Tighten any loose item(s).
2.
Check that all controls and connectors are free of dust.
3.
Check that all O-rings in the headplate are intact and in good condition. Replace
those that are not.
18.5 Fuse
Replacement
There is one replaceable 5-Amp fast-acting glass tube fuse.
18.6
Port & Adapter O-Rings
Each time you clean the vessel assembly, check all port and adapter O-rings for wear,
distortion, rips or any other damage to their integrity. Replace them as needed.
#
NOTE:
Only hand tighten ports: never use a wrench!
If you need to seal a leak, replace the O-ring.
18.7
Headplate O-Ring Replacement
When it needs to be replaced:
1.
Pull the old O-ring out with a knife or a flat-bladed screwdriver.
2.
Lubricate the new O-ring with a light coat of silicone grease.
3.
Insert the new O-ring, making sure it is well-seated in the groove.
Содержание BioFlo 415
Страница 12: ......
Страница 128: ...116 BioFlo 415 y M1360 0050 User s Guide 2 20 0 D DR RA AW WI IN NG GS S Figure 65 Controller Schematics overview ...
Страница 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Страница 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Страница 145: ...133 New Brunswick Scientific User s Guide ...
Страница 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Страница 147: ...135 New Brunswick Scientific User s Guide ...
Страница 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Страница 149: ...137 New Brunswick Scientific User s Guide ...
Страница 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Страница 151: ...139 New Brunswick Scientific User s Guide ...
Страница 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Страница 153: ...141 New Brunswick Scientific User s Guide ...
Страница 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Страница 155: ...143 New Brunswick Scientific User s Guide ...
Страница 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Страница 157: ...145 New Brunswick Scientific User s Guide ...
Страница 158: ...146 BioFlo 415 y M1360 0050 User s Guide ...
Страница 164: ...152 BioFlo 415 y M1360 0050 User s Guide ...