Figure 29: Calibration Screen (DO)
Method 2:
#
NOTE:
Nitrogen is needed for Method 2. There is an N
2
gas inlet (“GAS-3”) on
the control cabinet for this purpose; make sure that your nitrogen source
is connected to this inlet.
1.
Connect the DO cable to the DO electrode and the control cabinet.
2.
Go to the
CALIBRATION
screen and select
DO
(
see Figure 29
).
3.
Press the
N2 (3) ON
button. If your system has 3 or 4 TMFCs, however,
this button will not be present. In this case, manually turn the N
2
loop on
from the
SUMMARY
screen and set it to 1-20 SLPM (depending on vessel
size and flow controller).
4.
In approximately 10-30 minutes, the current value reading will stabilize.
5.
Press the
SET ZERO
edit box, use the touchpad to enter
0
, press the
OK
button, then press the
SET ZERO
button.
6.
Press
N2 (3) OFF
(or, if in Step 3 you manually turned the N
2
loop on, now
manually shut off the nitrogen flow to the vessel).
7.3.6
DO Probe Calibration: Setting Span
1.
In the
AGIT GAUGE
screen, set the
AGIT
speed to
50 RPM.
2.
Set the
AGIT
mode to
AUTO
.
3.
Vigorously sparge air into the vessel via the filter on the headplate until the
display is stable for approximately 10 minutes (this may take up to 30
minutes total).
4.
In the
CALIBRATION
screen, select
DO
(
see Figure 29
).
5.
Enter
100
in the
SET SPAN
edit box, then press the
SET SPAN
button.
2. Open the
CALIBRATION
screen and
press DO.
3. Enter 0 in
the SET ZERO
edit box.
4. Press the SET
ZERO button.
5. Reconnect the
DO cable to the
DO electrode.
Содержание BioFlo 415
Страница 12: ......
Страница 128: ...116 BioFlo 415 y M1360 0050 User s Guide 2 20 0 D DR RA AW WI IN NG GS S Figure 65 Controller Schematics overview ...
Страница 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Страница 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Страница 145: ...133 New Brunswick Scientific User s Guide ...
Страница 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Страница 147: ...135 New Brunswick Scientific User s Guide ...
Страница 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Страница 149: ...137 New Brunswick Scientific User s Guide ...
Страница 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Страница 151: ...139 New Brunswick Scientific User s Guide ...
Страница 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Страница 153: ...141 New Brunswick Scientific User s Guide ...
Страница 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Страница 155: ...143 New Brunswick Scientific User s Guide ...
Страница 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Страница 157: ...145 New Brunswick Scientific User s Guide ...
Страница 158: ...146 BioFlo 415 y M1360 0050 User s Guide ...
Страница 164: ...152 BioFlo 415 y M1360 0050 User s Guide ...