Specifications
88
Patient Risk Applied Current
Patient Isolation Risk Current
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1,
UL 2601
-
1
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
100
µ
A
10
µ
A
Normal
Open
Closed
500
µ
A
50
µ
A
Normal
Closed
Open
500
µ
A
50
µ
A
Reversed
Closed
Closed
100
µ
A
10
µ
A
Reversed
Open
Closed
500
µ
A
50
µ
A
Reversed
Closed
Open
500
µ
A
50
µ
A
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1,
UL 2601
-
1
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
5 mA
50
µ
A
Reversed
Closed
Closed
5 mA
50
µ
A