Troubleshooting
N-550
47
Display/Alarms
Table 8 lists symptoms of problems relating to non-functioning displays and
audible tones or alarms and recommended actions. If the action requires
replacement of a PCB or module, refer to
Table 8: Display/Alarms Problems
Symptom
Recommended Action
Display values are
missing or erratic.
Try another sensor or relocate the sensor.
Replace the Sensor Extension Cable.
Replace the sensor.
Replace the Front Panel PCB. See
Not all display segments
light during POST.
Check the connection between the Front Panel PCB
and the Main PCB.
Replace the Front Panel PCB. See
All Front Panel LED
indicators do not light
during POST
Check the connection between the Front Panel PCB
and the Main PCB.
Replace the front case. See
Alarm sounds for no
apparent reason.
Moisture or spilled liquid can cause an alarm to sound.
Allow the N-550 to dry thoroughly before use.
Replace the Front Panel PCB as described in
Display is flashing, but
there is no alarm.
Verify that alarm silence has not been activated. See
Check speaker connection to Front Panel PCB. Refer
to
Replace the speaker. Refer to
Replace the Front Panel PCB. Refer to
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...