Disassembly Guide
N-550
61
MP-506 Removal
1. Perform the procedure
2. Perform the procedure
Separating Top and Bottom Cases
3. Disconnect the front case SpO
2
cable connector from the MP-506 PCB
connector. See Figure 22 item 1.
Figure 22: MP-506 Cable Connections
4. Disconnect the main PCB cable connector from the MP-506 PCB
connector, item 2.
5. Carefully lift the MP-506 PCB out of the top case. See Figure 23.