A u d i b l e A l a r m S e t t i n g s a n d S e r v i c e
F u n c t i o n s
Introduction
This section discusses use of the service menu to reconfigure power-on
default values, and how to control the behavior of the audible alarm.
Audible Alarm Settings
The following paragraphs describe how to change the behavior of the audible
alarm. Operators can select the volume of the alarm and the duration of alarm
silence. Controls for the N-550 are shown in Figure 9.
Figure 9: N-550 Controls
Alarm Silence State
Press the Alarm Silence button to silence the alarm. Press the Alarm Silence
button a second time to turn the alarm back on.
1 — Power On/Off Button
5 — Pulse Rate Alarm Limit Button
2 — Alarm Silence Button
6 —
SatSeconds
Alarm Limit Button
3 — Adjust Up Button
7 — SpO
2
Alarm Limit Button
4 — Adjust Down Button