Audible Alarm Settings and Service Functions
30
Alarm Silence Duration
1. Press and hold the Alarm Silence button.
2. Press the Adjust Up or Adjust Down button to change the duration of the
alarm silence. The alarm duration can be set to 30, 60, 90, or 120 seconds,
or the alarm can be turned to Off.
Alarm Volume
1. Press and hold the Alarm Silence button until “VOL” is displayed in the
%SpO
2
window.
2. Press the Adjust Up or Adjust Down button to select alarm volumes from
1 to 10. Select a level that is suitable for the N-550's location.
N-550 Menu
Table 2 lists the menu selections available on the N-550. Items in
BOLD
are
the default settings.
Table 2: N-550 Menu Selections
Menu
Number
Sub-Menu
Number
Function
1
-
Trend print (tabular N-550 trend only)
2
-
Clear N-550 trend
3
0
Language
English
4
-
Data port baud rate
24
2400 bauds per second (bps)
96
9600 bps
192
19200 bps
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...