Performance Verification
N-550
9
Figure 3: N-550 Controls
Power-On Self-Test (POST)
CAUTION: If any indicator or display element does not light, or the
speaker does not sound, do not use the N-550. Instead, contact qualified
service personnel, your local Nellcor representative, or Nellcor's
Technical Services Department, 1.800.635.5267.
Note:
The N-550 should complete the POST function within 12 seconds.
1. Turn on the N-550 by pressing the Power On/Off button for greater than
0.5 seconds.
2. The N-550 displays/sounds:
1 — Power On/Off Button
5 — Pulse Rate Alarm Limit Button
2 — Alarm Silence Button
6 —
SatSeconds
Alarm Limit Button
3 — Adjust Up Button
7 — SpO
2
Alarm Limit Button
4 — Adjust Down Button
Display
Sound
5 (in pulse rate left window)
low priority alarm
5 (in pulse rate center window)
medium priority alarm
0 (in pulse rate right window)
high priority alarm
n (in SpO
2
left window)
n (in SpO
2
center window)
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...