Audible Alarm Settings and Service Functions
N-550
41
When the desired option is indicated in the %SpO
2
display, press the
SatSeconds
Alarm Limit button to set the current selection. The N-550 sounds
a confirmation tone.
Setting Institutional Defaults (Sample)
Set alarm limits to the values established by your facility. After all alarm
limits are set, you will access menu Item 9 to set the alarm limit values to
institutional defaults. The following default values may be set:
•
Alarm Suspend (30, 60, 90, 120 seconds)
•
Alarm Silence Behavior (silence off, sound reminder, do not allow
OFF)
•
Alarm Volume (1 to 10)
•
Baud Rate (2400, 9600, 19200)
•
Data Port Mode (ASCII)
•
Pulse Beep Volume (0 to 10)
•
Pulse Rate Upper Alarm Limit (lower limit plus one to 250 bpm)
•
Pulse Rate Lower Alarm Limit (30 bpm to upper limit minus one)
•
SatSeconds
(OFF, 10, 25, 50, 100)
•
SpO
2
Upper Alarm Limit (lower limit plus one to 100%)
•
SpO
2
Lower Limit (20% to upper limit minus one)
1. Disconnect the sensor from the N-550.
2. Press the Power On/Off button to turn the N-550 off.
3. Press the Power On/Off button to turn the N-550 on.
4. Set the desired parameters to the institutional values.
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...