Performance Verification
14
12. Press the Pulse Rate Alarm Limit button. Verify that the N-550 emits a
single beep, the pulse rate display indicates an alarm limit of “170” and
that the %SpO
2
display window shows “HI” for about 3 seconds.
Figure 6: Adjusting High Pulse Rate Alarm Limit
13. Press the Pulse Rate Alarm Limit Button. Within 3 seconds press and hold
the Adjust Down button. Verify that the pulse rate display reduces to a
minimum of one point above the pulse rate low limit alarm setting.
14. Press the Pulse Rate Alarm Limit button two times (twice within 3
seconds). Verify that the N-550 emits two beeps and that the pulse rate
display indicates an alarm limit of “40” for 3 seconds. Verify that “LO” is
displayed in the %SpO
2
display.
Figure 7: Adjusting Low Pulse Rate Alarm Limit
15. Press the Pulse Rate Alarm Limit button two times rapidly. Within 3
seconds press and hold the Adjust Down button. Verify that the pulse rate
display cannot be raised past the upper alarm limit setting minus one
point.
16. Turn the N-550 off by pressing the Power On/Off button.
17. Turn the N-550 on by pressing the Power On/Off button.
Содержание OxiMax N-550
Страница 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Страница 10: ...Blank Page ...
Страница 38: ...Blank Page ...
Страница 64: ...Blank Page ...
Страница 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Страница 82: ...Blank Page ...
Страница 86: ...Spare Parts 76 Figure 40 Exploded View ...
Страница 90: ...Blank Page ...
Страница 108: ...Blank Page ...
Страница 120: ...Blank Page ...
Страница 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Страница 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Страница 136: ...Blank Page ...
Страница 137: ... ...