Chapter 1 Overview
1.6 Connecting The Electrode and the Power Source
F-52, 53, 54, 55
13
Electrode connector
Connect the measurement electrode to CH1 (or CH2).
The connector cover functions to protect the connector from water to be splashed. Be
sure to attach it to the connector.
1.
Match the groove on the electrode
connector with the pin on the
counterpart located on the main
unit.
Do not push the cover to the
connector forcefully when the slit
does not fit to the pin.
Do not work on the connector with dirty or
wet hands.
2.
Hold the metal part of the electrode
connector, and push it according to
the slit by rotating clockwise.
Pinched injury
3.
Cover the connector with the con-
nector cover on the straight.
Never rotate the connector cover. It will be
blown.
Temperature connector
1.
According to the connection CH of
the measurement electrode, con-
nect the temperature connector.
2.
Fully insert the temperature con-
nector into the counterpart located
on the main unit.
If measurement is performed in the auto-
matic temperature compensation (ATC)
mode without connecting the temperature
connector or if the mV measurement is car-
ried out in that state, the temperature reading
will be always 25°C.
Connector cover
Temperature
electrode
Содержание F-52
Страница 2: ......
Страница 13: ...TABLE OF CONTENTS X HORIBA...
Страница 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Страница 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Страница 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Страница 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Страница 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Страница 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Страница 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...