Chapter 3 pH MEASUREMENT
3.4 Calibration Setting
48
HORIBA
3.4.5 Preliminary check setting
By confirming the repeatability using the standard solution of pH7 once a day at calibration,
the measurement accuracy can be assured.
When calibration for the set calibration point is completed, the repeatability is automatically
checked. This repeatability check confirms the deviation from the calibrated value by the
standard solution of pH7.
The sequence for the two-point calibration by pH7 and pH4 is as follows:
pH7 (Calibration of 1st point)
→
pH4 (Calibration of 2nd point)
→
pH7(Repeatability check
with the calibrated value by 1st point )
The result is displayed as follows:
Preliminary check is only available for NIST and USA in standard solution setting.
It is not available for Custom setting.
Setting procedure
1.
In the pH CALIBRATION SETUP
screen, press
↑
or
↓
key to go to
REPEATABILITY field.
2.
Press
←
or
→
key to select ON or
OFF.
3.
Press F4 (RETURN) to reflect the
setting. The display returns to the
CAL screen.
Screen display
Repeatability
OK(0)
Within
±
0.005 pH
OK(1)
Within
±
0.02 pH
OK(2)
Within
±
0.05 pH
Error
Over
±
0.05 pH
Содержание F-52
Страница 2: ......
Страница 13: ...TABLE OF CONTENTS X HORIBA...
Страница 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Страница 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Страница 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Страница 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Страница 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Страница 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Страница 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...