Chapter 12 RS-232C Communication
12.4 List of Serial Command
F-52, 53, 54, 55
161
●
Data request command (2)
Format
Header, Item code, Parameter <CR><LF>
Command
for...
Command
Parameter
Parameter
description
Response
Note
Header
Comma
nd code
Ion calibra-
tion history
request
R
IC
x
x: channel number
RIC,zzzzzzzzzzzz,
x,aa,b,c,d,eeeeee,f
,yyyy,yy,yy,yy,yy,yy,
pppp,q,r,sssss,tttttt
t,uuuuuu,
…
Response when
there is no calibra-
tion data
RIC,************,x,
0,0,3
zzzzzzzzzzzz:
Operator name
x: channel number
aa: ion type
b: number of cali-
brated points
c: Electrode status
d: Temperature com-
pensation
eeeeee:
Asymmetry potential
f: Check flag before
use
yyyy
…
yy:
Calibration date
pppp: Calibration solu-
tion value
q: Sub unit
r: Unit
sssss: temperature
ttttttt: Potential
uuuuuu: Sensibility
Repeats calibration
points only after cali-
bration date.
Check data before
use is added if any.
Clock data
request
R
OT
ROT,yyyy,mm,dd,h
h,nn,ss
yyyy: Year
mm: month
dd: Day
hh: Hours
nn: minutes
ss: Seconds
Содержание F-52
Страница 2: ......
Страница 13: ...TABLE OF CONTENTS X HORIBA...
Страница 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Страница 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Страница 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Страница 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Страница 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Страница 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Страница 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...