Chapter 3 pH MEASUREMENT
3.4 Calibration Setting
F-52, 53, 54, 55
55
The measurement values during the periodical check are displayed after the calibra-
tion is cleared.
4.
When the check is completed, the result data is automatically printed out.
●
OK/CHECK
OK (satisfies the regulation)/CHECK
(does not satisfy the regulation)
●
Calibration
●
Asymmetry potential
(Within regulated value
±
3 mV)
●
Sensitivity
(99% to 101% of regulated value)
●
Linearity
Indication error at each input
(Regulated value
±
0.010 pH
±
1digit)
●
High impedance
Indication error at each input
(Regulated value
±
0.030 pH
±
1digit)
●
Temperature
Indication error at each input
(Regulated value
±
0.4
℃
)
Sample printout
When the judgment of “CHECK” is displayed, check the connection and repeat the check.
When the error repeatedly occurs, call customer service.
OK/CHECK judgment
Date of check
Linearity
Calibration by standard solution
Temperature