Chapter 12 RS-232C Communication
12.4 List of Serial Command
156
HORIBA
●
On-Line Operation Commands(1)
Format
Header, Item code, Parameter <CR><LF>
In RS-232C communication, the relative mV measurement is not available.
Command
for...
Command
Parameter
Description of
parameter
Note
Header
Comma
nd code
On-line/
off-line
C
OL
x
0: On-line
1: Off-line
ON
→
ON, OFF
→
OFF: The state
remains unchanged.
OFF
→
ON:
Proceeds to standby mode
for measurement in the main screen.
ON
→
OFF: No screen transition
Effective when the screen is in Mea-
surement. For other cases, the error
response (malfunction) returns.
Halt
potential
hunting
C
BR
x
x: channel number
The measurement or calibration in Auto-
Hold is aborted.
In the measurement with AutoHold OFF
or in calibration, either calibration value
is forcefully obtained (in calibration), or
holds the measured value (in measure-
ment).
In the interval memory, the error
response (malfunction) returns.
pH measure-
ment mode
specification
C
PH
x
x: channel number
pH Measurement proceeds to standby
mode for the specified channel.
mV measure-
ment mode
specification
command
C
MV
x
x: channel number
Falls into the measurement ready condi-
tion in the ABS. ORP measurement
mode in the specified channel.
Ion measure-
ment mode
specification
C
IO
x
x: channel number
Falls into the ion measurement ready
condition in the specified channel.
Conductivity
measure-
ment mode
specification
C
CO
Falls into the COND measurement
ready condition.
Saline mea-
surement
specification
C
SA
Falls into the salinity measurement
ready condition.
Resistivity
measure-
ment specifi-
cation
C
OH
Proceeds to standby mode for resistivity
measurement.
Содержание F-52
Страница 2: ......
Страница 13: ...TABLE OF CONTENTS X HORIBA...
Страница 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Страница 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Страница 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Страница 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Страница 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Страница 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Страница 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...