Chapter 16 Reference
16.5 Specifications
246
HORIBA
Resis-
tivity
Measuring
principle
●
●
Conversion from conductivity value
Measuring
range
●
●
Cell constant
100 m
-1
:
5.00
Ω
m to 199.9 k
Ω
•m
Cell constant
10 m
-1
:
50.0
Ω
m to 1.999 M
Ω
•m
Cell constant
1000 m
-1
:
0.500
Ω
m to 19.99 k
Ω
•m
Resolution
●
●
0.05% of full scale
Repeatability
●
●
±0.5% ±1 digit of full scale
Input
CH number
●
1CH
●
●
●
2CH
Output RS-232C
●
●
●
●
Printer output
●
●
●
●
Analog output
●
●
●
●
Compact flash
●
●
●
Ethernet
output
●
Data
mem-
ory
Memory
number
●
●
●
●
300 pcs
Interval
memory
●
●
●
●
Dis-
play
Display
●
●
●
●
Color graphic LCD
Navigation
function
●
●
●
●
Operator
registration
●
●
●
●
Security
function
●
●
●
●
AutoHold
setting
●
●
●
●
pH
number of cali-
brated points
●
●
●
●
5
Repeatability
confirmation
●
●
●
●
Periodical
inspection
●
●
●
●
ION
Calibration
curve method
●
●
5
Addition
method
measurement
●
●
COND Distilled water
temperature
conversion
●
●
F-52 F-53 F-54 F-55
Description
Содержание F-52
Страница 2: ......
Страница 13: ...TABLE OF CONTENTS X HORIBA...
Страница 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Страница 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Страница 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Страница 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Страница 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Страница 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Страница 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...