Chapter 15 Maintenance and troubleshooting
15.2 Troubleshooting
F-52, 53, 54, 55
205
●
ERR No.08 Calibration cycle error
Definition
This error appears when the number of mea-
surements set for the calibration cycle has
been exceeded since the last calibration was
conducted.
Perform calibration.
●
ERR No.09 There is a problem with the printer unit
Definition
If a problem occurs with the printer unit, turn
OFF the power to the meter, perform the
appropriate measure below, then turn the
power to the meter back ON.
●
ERR No.10 Data memory over
Definition
The number of data items has exceeded the
limit of the memory.
Cause
How to solve problem
The printer paper is
jammed.
Remove the jammed paper.
There is no printer
paper.
Load new paper in the printer.
There is a problem
with the printer unit
connection.
Reconnect the printer after making sure the
connector parts are okay.
The printer is defec-
tive.
Replace the printer.
Contact with the nearest sales representative
when the function is not restored even after
the printer is replaced.
Cause
How to solve problem
Memory over
Delete data stored in the memory confirming
their contents.
Содержание F-52
Страница 2: ......
Страница 13: ...TABLE OF CONTENTS X HORIBA...
Страница 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Страница 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Страница 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Страница 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Страница 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Страница 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Страница 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...