Chapter 11 METER SETTING
11.4 Maintenance Setting
F-52, 53, 54, 55
143
Temperature sensor calibration for CH1
The indication value for the temperature sensor connected to CH1 can be adjusted.
1.
Set each temperature compensation setting to “Automatic.”
2.
Immerse the temperature sensor into a solution which temperature has been
stabilized.
3.
Read the temperature indication correctly by a thermometer.
4.
Press the ENTER key.
The setting is now ready to be made.
5.
Press
↑
or
↓
key and set the value to the read temperature from the thermom-
eter.
6.
Press the ENTER key.
The setting is made.
When the temperature sensor is not connected, this function does not work properly.
Temperature sensor calibration for CH2
(
F-53,54,55
)
The indication value for the temperature sensor connected to CH2 can be adjusted.
For operation, refer to the above, “Temperature sensor calibration for CH1”.