Chapter 11 METER SETTING
11.4 Maintenance Setting
F-52, 53, 54, 55
141
Analog test
Analog output voltage can be checked and
adjusted.
1.
Press the ENTER key.
The ANALOG OUTPUT CHECK
screen appears.
2.
Press
↑
or
↓
key and select the
item, and press ENTER key to
determine the selection.
3.
Check the output by a voltmeter.
4.
Press
↑
or
↓
key, and adjust the
output voltage indicated on the
voltmeter to the specified values,
0V/2V.
5.
Press the ENTER key to assure
the output.
6.
When the confirmation and/or
adjustment is completed, press F1
(CANCEL).
The display returns to the MAINTE-
NANCE screen.